Query :MACHine{1|2}:STRigger:RANGe<N>?
The RANGe query returns the range recognizer end point specifications for
the range.
Returned Format [:MACHine{1|2}:STRigger:RANGe<N>]
<label_name>,<start_pattern>,<stop_pattern><NL>
Example OUTPUT XXX;":MACHINE1:STRIGGER:RANGE1?"
SEQuence
Command :MACHine{1|2}:STRigger:SEQuence <num_levels>,
<trig_level>
The SEQuence command redefines the state analyzer trigger sequence.
First, it deletes the current sequence. Then it inserts the number of levels
specified, with default settings, and assigns the trigger to be at a specified
sequence level. The number of levels may be between 2 and 12.
<num_levels> integer from 2 to 12
<trig_level>
integer from 1 to (number of existing sequence levels 1)
Example OUTPUT XXX;":MACHINE1:STRIGGER:SEQUENCE 4,3"
Query :MACHine{1|2}:STRigger:SEQuence?
The SEQuence query returns the current sequence specification.
Returned Format
[:MACHine{1|2}:STRigger:SEQuence] <num_levels>,
<trig_level><NL>
Example OUTPUT XXX;":MACHINE1:STRIGGER:SEQUENCE?"
STRigger (STRace) Subsystem