5-10
HP 8517B S-Parameter Test Set Manual
Troubleshooting the Test Set
Troubleshooting Procedures
Figure 5-6 The ACTIVE LED Location
If the Self-Test Fails to Run Properly
If the portion of memory holding the self-test program is faulty, the self-test
will not run properly. If any of the following conditions occur, it indicates
thattheA4HP-IBboardislikelytobefaulty.
All LEDs flash briefly, then turn off
All LEDs flash briefly and stay on
The ACTIVE LED turns on too quickly after power on
The ACTIVE LED does not turn on at all