Series 2040 Test System
Series 2040 Maintenance Manual V2.00Selftest & Calibration
10
Utilities” dialog box in shown on the
next page will appear. This box contains
three separate ways of reporting the
data collected during the tests. Data
may be stored to a file, displayed on the
screen and/or printed.
In addition, each of these three
reporting functions can report “All
Data”, “Fail Data” or “No Data”. “All
Data” includes all of the pass or fail
sequences from the test. “Fail Data
includes only information on the tests
that failed. “No Data” disables that reporting function. To select, merely
use the mouse to check or uncheck each reporting function. If a
reporting device is unchecked, the options under that device appear
ghosted”, indicating that they are disabled. For diagnostic purposes,
enable the “Printer” and “File” options, and select the “Fail Data
option. Under the “File” section, click the “Diagnostic File” check box.
Disable the Window reporting functions.