Chapter 4 Command Reference
Introduction
Programming Guide 4-15
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
:CALCulate2:LIMit:FCOunt:LOWer?
Queries the number of limit test failures (that is, the Fail COunt) at the lower
limit.
Numeric data transferred as ASCII bytes in <NR1> format.
If CALC2:LIM:STATe is OFF, 0 is returned and error -221 is generated.
If no valid result exists, 0 is returned and error -230 is generated.
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
Query only.
:CALCulate2:LIMit:FCOunt[:TOTal]?
Queries the total Fail COunt (that is, the number of measurements that have failed
the limit test). No failures is indicated by 0.
Numeric data transferred as ASCII bytes in <NR1> format.
If CALC2:LIM:STATe is OFF, 0 is returned and error -221 is generated.
If no valid result exists, 0 is returned and error -230 is generated.
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
Query only.
:CALCulate2:LIMit:FCOunt:UPPer?
Queries the number of limit test failures (that is, the Fail COunt) at the upper
limit.
Numeric data transferred as ASCII bytes in <NR1> format.
If CALC2:LIM:STATe is OFF, 0 is returned and error -221 is generated.
If no valid result exists, 0 is returned and error -230 is generated.
If the current measurement is Totalize or Voltage Peaks, 0 is returned and
error -221 is generated.
Query Response
Comments
Query Response
Comments
Query Response