HP 132A Manual do Utilizador Página 202

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 330
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 201
Chapter 4 Command Reference
:DIAGnostic Subsystem
4-36 Programming Guide
:DIAGnostic:CALibration:TINTerval:QUICk
This event command calibrates out the differences in electrical path length
between channels 1 and 2.
Before sending this command, connect to channel 1 a square wave of approximate
frequency 10 MHz, but more importantly with a rapid rise time.
No query.
The calibration values are stored in non-volatile memory, and are unaffected
by power-on, save/recall, and *RST.
Use DIAG:CAL:STAT? to check for successful calibration.
Scale and Offset/POWER (Calibration menu).
Comments
Related
Front-Panel
Keys
Vista de página 201
1 2 ... 197 198 199 200 201 202 203 204 205 206 207 ... 329 330

Comentários a estes Manuais

Sem comentários