HP Waterskis E1312A Manual do Utilizador Página 97

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 212
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 96
Multimeter Command Reference 97Chapter 3
DATA
The multimeter can store up to 512 readings in internal memory. The DATA
command allows you to determine how many readings are currently stored.
Subsystem Syntax DATA
:POINts?
:POINts?
The INITiate command uses internal memory to store readings prior to a FETCh?
command e.g., when a measurement is initiated by the
INITiate command. You can
query the number of stored readings in memory by sending the
DATA:POINts?
command.
Comments INITiate command uses internal memory to store readings prior to using a
FETCh? command. You use the DATA:POINts? command to query the
number of readings stored in internal memory to determine the amount of data
space to allocate on your computer to receive the data.