I
-
I
-
Using Spectrum Analyzer Features
learn How to Enter Amplitude Correction Factors
Q
3
Q
4
14:85:24
AUG 18,
1992
I
I
Figure 3-13. Amplitude-Correction Points
Item
Description of Items in Figure 3-13
1 Frequencies below first point use first amplitude level.
2
First segment interpolated with the 10
dB
amplitude level.
3
Frequency and amplitude coordinate that starts the second segment.
4
Third segment interpolated with the -10
dB
amplitude level.
5
Frequency and amplitude coordinate that starts the fourth segment.