HP 8591C Manual do Utilizador Página 194

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 550
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 193
I
-
Making Basic Measurements Using Spectrum Analyzer Mode
Identifying Distortion Products Using the RF Attenuator and Traces
If the responses in trace A and trace B differ, as in Figure 4-2
1,
then
attenuation is required. If there is no change in the signal level, the distortion
is not caused internally. For example, the signal amplitude in Figure 4-22 is
not high enough to cause internal distortion in the spectrum analyzer so any
distortion that is displayed is present on the input signal
LOG
*
10
I
I I
I I I I I I
SCALE
LOG
LIN
INT
AMP
ON
OFF
WA
“B
SC
FS
CORR
More
1
Of
2
I
CENTER
400.0
MHZ
SPAN
200.0
MHZ
RES
BW
1
.o
MHZ
“EW
300
kHZ
SWP
20.0
m!sec
Figure 4.22. No Harmonic Distortion
Vista de página 193
1 2 ... 189 190 191 192 193 194 195 196 197 198 199 ... 549 550

Comentários a estes Manuais

Sem comentários