I
-
Making Measurements Using Spectrum Analyzer Mode
Using the Time-Gated Spectrum Analyzer Capability Without the Gate Utility (Option 107 only)
3.
Press
Prevfous
Menu, GATE
ON
OFF (so that ON is underlined). See
Figure
5-30.
NOTE
This procedure offers a qualitative functional check only Due to several factors, the accuracy of
the marker readout of the gate delay and gate length can vary
by
several milliseconds. For more
information about gate timing, see the Characteristics in specifications and characteristics in your
calibration guide for your spectrum analyzer.
L
,
Figure
5-30.
Gate On
Item
Description of Items in Figure
5-30
1
Represents the gate delay. The gate is
off
during the gate delay.
2
Represents the gate length. The gate is on, and the HI SWEEP IN/OUT signal is displayed.
5-57