HP 34970A Manual do Utilizador Página 213

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 212
[SENSe:]
VOLTage:DC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
VOLTage:AC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
RESistance:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
FRESistance:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
CURRent:DC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
CURRent:AC:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
FREQuency:VOLTage:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
PERiod:VOLTage:RANGe {<
range>|MIN|MAX}[,(@<ch_list>)]
Select the measurement range for the function selected on the specified
channels.
MIN selects the lowest range. MAX selects the highest range.
[SENSe:]
VOLTage:DC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
VOLTage:AC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
RESistance:RANGe? [{(
@<ch_list>)|MIN|MAX}]
FRESistance:RANGe? [{(
@<ch_list>)|MIN|MAX}]
CURRent:DC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
CURRent:AC:RANGe? [{(
@<ch_list>)|MIN|MAX}]
FREQuency:VOLTage:RANGe? [{(
@<ch_ list>)|MIN|MAX}]
PERiod:VOLTage:RANGe? [{(
@<ch_list>)|MIN|MAX}]
Query the measurement range on the specified channels. Returns a
number in the form “
+1.00000000E+01”.
[SENSe:]
VOLTage:DC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
VOLTage:AC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
RESistance:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
FRESistance:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
CURRent:DC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
CURRent:AC:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
FREQuency:VOLTage:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
PERiod:VOLTage:RANGe:AUTO {OFF|ON}[,(
@<ch_list>)]
Disable or enable autoranging on the specified channels. Autoranging
thresholds: Down ranges at <10% of range; Up range at >120% of range.
5
Chapter 5 Remote Interface Reference
Setting the Function, Range, and Resolution
215
Vista de página 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 428 429

Comentários a estes Manuais