[SENSe:]
VOLTage:DC:APERture {<
time>|MIN|MAX}[,(@<ch_list>)]
RESistance:APERture {<
time>|MIN|MAX}[,(@<ch_list>)]
FRESistance:APERture {<
time>|MIN|MAX}[,(@<ch_list>)]
CURRent:DC:APERture {<
time>|MIN|MAX}[,(@<ch_list>)]
Select the aperture time for the function selected on the specified channels.
MIN selects the smallest value accepted for this parameter, which gives
the most resolution.
MAX selects the largest value accepted for this
parameter, which gives the least resolution.
For more information on aperture time, refer to “Custom A/D Integration
Time” in chapter 4 starting on page 103.
[SENSe:]
VOLTage:DC:APERture? [{(
@<ch_list>)|MIN|MAX}]
RESistance:APERture? [{(
@<ch_list>)|MIN|MAX}]
FRESistance:APERture? [{(
@<ch_list>)|MIN|MAX}]
CURRent:DC:APERture? [{(
@<ch_list>)|MIN|MAX}]
Query the aperture time setting on the specified channels. Returns a
number in the form “
+1.66666700E-02”.
[SENSe:]
FREQuency:APERture {0.01|0.1|1|MIN|MAX}[,(
@<ch_list>)]
PERiod:APERture {0.01|0.1|1|MIN|MAX}[,(
@<ch_list>)]
Select the aperture time (or gate time) for frequency or period
measurements on the specified channels. Specify 10 ms (4
1
2
digits),
100 ms (default; 5
1
2
digits), or 1 second (6
1
2
digits). MIN = 0.01 seconds.
MAX = 1 second.
[SENSe:]
FREQuency:APERture? [{(
@<ch_list >)|MIN|MAX}]
PERiod:APERture? [{(
@<ch_list>)|MIN|MAX}]
Query the aperture time for frequency or period measurements on the
specified channels. Returns a number in the form “
+1.00000000E-01”.