Chapter 5 Examining Spectrum Data
5-54 Applied Biosystems
5
General
guidelines for
setting
parameters
Refer to the following table to determine how to set Advance
Baseline Correction parameters and obtain the desired
baseline correction.
Condition Set Parameters To
Baseline follows a rising peak cluster Peak WidthAverage peak width
Flexibility0.5
Degree0.5
Baseline is gently curving Peak Width10 to 20 times peak width
Flexibility0
Degree0.5 (decrease further to raise
the baseline)
Baseline rise should be ignored and
treated as signal
Peak WidthHalf of the width of the
cluster to correct (in the illustration
shown, if the width of the cluster is 200,
set Peak Width to 100)
Flexibility0
Degree0.5 (decrease further to raise
the baseline)
Peak cluster