Chapter 8 Viewing Voyager PSD Data
8-16 Applied Biosystems
8
Selecting peaks
manually
For optimum mass accuracy, select peaks as described in
Selecting calibration peaks for optimum mass accuracy on
page 8-19.
To manually select the reference mass for a peak:
1. Right-click-drag over the peak of interest.
The Reference Mass dialog box (Figure 8-6) is
displayed and lists all masses in the calibration
reference file. The entry highlighted is the nearest
match in the calibration reference file to the selected
peak that is within the mass tolerance specified.
Figure 8-6 Reference Mass Dialog Box