Calibration11-2
Pen-to-Paper Calibration (PPC)
Installation of PPC software
Step 1. Install the PPC software by running the set up file. Please accept all the default
settings. The software is self installing without much help.
Step 2. Using Windows Explorer, go to c:\ppc. Create a shortcut to file, ppc.exe, at the
desk top.
Step 3. Double click on the ‘ppc icon’. A login menu will appear and give the following
details.
Login :- hp2000
Password :- rogue1
Step 4. Click the OK button when the software screen appears.
Step 5. Go to the ‘set up’ menu at the top of the screen. Select ‘Calibration stand’. A
detailed menu on Calibration stand appears. Key in all the details as per the
Calibration stand supplied along with the serial number. Press OK.
Step 6. Go to ‘set up’ menu. Select ‘Spec limits’ then ‘adjust’ menu. Change the
specifications as follows.
Min specification :- 1.1938
Nominal :- 1.3716
Max specification :- 1.5494
Let the other details remain unchanged as we will not be using the other details
for the measurement process. Press OK to continue.
Step 7. Click the left most button ‘CALIBRATION’. Put in the sensor serial number
and the calibration block serial number. Follow the menu to calibrate. There are
two sections for calibration. After the first stage, it will once again verify the
calibration.
Step 8. Press the big ‘MEASURE’ button on the right side of the screen. You can see
that the tool is ready for calibration.