place. Try Test box calibration using the reference microphone of the on-ear probe microphone in place of
the test box reference microphone. If this is fails, the on-ear probe microphone is defective and must be
replaced.
Invalid on-ear calibration
Invalid on-ear microphone calibration!
This message may be displayed during the on-ear calibration procedure. It indicates that the calibration curve is
significantly different from what is expected. It may be accompanied by an on-ear reference microphone error.
Check that a probe microphone assembly is securely plugged into the PROBE socket to the left of the test
chamber and that the cable has not become damaged.
Try using a new probe tube. Ensure it is correctly attached and positioned.
Barcode scanner malfunction
If there is no red scanning beam when the trigger is pulled, tug on the cable where it enters the scanner. If it
comes loose easily, push it back in until it clicks in place.
If the scanning beam stays on when the trigger is released or if the barcode scanner fails to read known good
barcodes, try the following:
1. Unplug the scanner from the equipment, then reconnect it. If the problem persists…
2. Print this page on an external printer (see External display, printer, speaker) or from the electronic user’s
guide (see Electronic user’s guide).
3. If your barcode scanner has a PSC logo, scan these two barcodes, starting with the top one. This will reset the
scanner to its default settings.