RM500SL
®
User's Guide Version 3.12
March 2015
5. Hold the probe module about 18 in. (0.5 m) in front of the sound-field speaker.
6. Hold the microphone of a calibrated sound level meter next to the on-ear reference microphone as shown. Set
the sound level meter to C weighting and slow.
7. The sound level meter should read 70 ±2 dB.
8. Repeat with the frequency at 250 Hz.
9. To stop the measurement, highlight and <PICK> [Stop test].
The probe module should be replaced if it fails this test.
Max TM SPL setup
The Max TM SPL safety feature allows you to set a limit on the SPL, measured by the probe microphone, above
which a test terminates. The MAX TM SPL limit is shown on all on-ear measurement screens. The default limit
is 120 dB.