The speech-reception threshold (SRT) is attained
when the LTASS is at threshold (approximately - de-
pending on test material and the individual). These
scenarios are shown in Figures 10 - 12 which follow.
It should be noted that analysis methods which use
shorter time periods produce higher peak levels and
signicantly dierent speech envelopes. In order to
produce results that can be directly compared to mea-
sures of threshold (and UCL), the analysis time period
needs to approximate the integration time of the ear.
Although this varies with frequency and individuals,
a value between 100 - 200 ms is likely. The Verit and
RM500SL use a 128 ms analysis time period as an ap-
proximation because it also has considerable historic
support.
Because the spectrum of a speech signal varies with
time, it is necessary to average measurements over
several seconds to obtain a stable, repeatable result.
However, the eects of adjustments to hearing aid pa-
rameters need to be seen quickly in order to be useful
for optimization. To resolve these conicting needs,
speech signals are rst presented in a repeating loop
with a sound-eld equalization test prior to the start
of each passage. A running calculation of the short-
term speech envelope and average is performed on
2 - 3 seconds worth of data and is displayed in real
time so that the eects of adjustments to hearing aid
parameters are easily viewed. Pressing the Continue
key causes the passage to restart and run in its entire-
ty. The envelope and average are calculated over the
full passage to provide stable and repeatable data.
When using live speech as a test signal, a “Freeze
curve function is available to capture the short-term
spectra for examination and counseling purposes.
Figure 9: Example of speech envelope and LTASS
09/01/29 © Etymonic Design Incorporated, 20 Ludwig St., Dorchester, ON, Canada N0L 1G4
USA 800-265-2093 519-268-3313 FAX 519-268-3256 www.audioscan.com
9