RM500SL
®
User's Guide Version 3.12
March 2015
5. Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To stop the
measurement, highlight and <PICK> [Stop test]. You may change Level and Frequency without stopping the
test.
Sound level meter using on-ear microphones
The manual control feature enables the RM500SL to be used as a sound level meter to perform C-weighted, A-
weighted or 1/12th octave analysis of sound as measured by the probe and reference microphones. A and C-
weighted measurements are limited to frequencies above 100 Hz and the averaging time and update rate are 384
ms.
1. Refer to On-Ear Measures - Setup to prepare the system for use and for proper positioning of the client and
the probe tube.
2. Press <Tests>, then highlight and <PICK> [On-ear measures], then highlight and <PICK> [Manual control].
3. Highlight and <PICK> the Level window, then highlight and <PICK> 0 dB stimulus level from the drop-
down list.
4. Highlight and <PICK> the Filter window, then highlight and <PICK> 1/12th octave, A-weighted, or C-
weighted from the drop-down list.
5. For 1/12th octave filter, highlight and <PICK> the Frequency window, then highlight and <PICK> the
desired frequency from the drop-down list.
6. Highlight and <PICK> [Start test] to start the test. [Start test] changes to [Stop Test]. To stop the
measurement, highlight and <PICK> [Stop test]. You may change Level and Frequency without stopping the
test.