RM500SL
®
User's Guide Version 3.12
March 2015
21 Troubleshooting
This section covers common problems encountered when using the RM500SL. If you require further assistance,
note your model and serial number (located next to the power switch) and contact your local distributor, or
Audioscan at 519-268-3313 800-265-2093 (USA only), 519-268-3256 (fax) or service@audioscan.com
PLEASE DO NOT SHIP YOUR UNIT BACK TO THE FACTORY. In most cases we can courier you a part
that you can quickly and easily replace yourself.
Self test failures
Any failure during power-on self test (P.O.S.T.) is indicated on-screen and further use of the instrument is
denied. Try the following steps to resolve the failure:
1. To access Self test, press <Setup>, then highlight and <PICK> [Self test].
2. If you received a Max out/in or a Routing failure, remove any hearing aids from the test chamber and rerun
self test.
3. Unplug all microphones (On-ear & Test box). Rerun self test. If self test passes, reconnect the microphones,
one by one, running self test each time until a failure occurs. If the failure returns, display and print the test
results and contact Audioscan for help.
Initialize Function
Initializing your Audioscan unit will reset it to default factory settings. This erases the user preference files
stored in internal memory. Any custom printer choice, network settings, test setups will need to be re-entered
and the user calibrations will also be required.
1. To initialize, press <Setup>, then highlight and <PICK> [Self test]
2. Highlight and <PICK> [Initilize]. Unit will automatically restart.
98